undefined
+
  • undefined

Animal-specific touch screen monitoring equipment Multiparameter

Key words:

Service Hotline:

  • Product Description
  •  

    Adopt international veterinary calculation standards, such as blood pressure measurement mode designed according to AAMI/ANSI SP10, '02. Animal types vary, micro-flow end-tidal carbon dioxide monitoring technology, etc.

    It is truly dedicated to animals and has the same principle as conventional human medical monitors. The difference is that CARDELL monitors use professional accessories suitable for animal characteristics and analysis software calculation methods suitable for animal characteristics, which greatly improves anti-interference technology and effectively avoids interference with signal acquisition caused by animal non-cooperation and hair and rough skin.

    Multiparameter digital high-definition animal monitor, Cardell veterinary blood pressure technology, Nellcor's fully digital Oximax animal pulse oximetry technology, Oridion's Microstream carbon dioxide technology. Using world-renowned veterinary diagnostic measurement technology, it is truly suitable for animals and can provide accurate animal electrocardiogram, blood oxygen, respiration, carbon dioxide and other parameter measurements.

    Compact, simple, reliable and durable physiological monitoring equipment; can greatly improve animal care; it is widely used in veterinary medicine, research institutions and hospitals, suitable for animals ranging from kittens to horses.

    parameter:

    Six routine parameters: ECG, heart rate HR, blood oxygen SPO2, non-invasive blood pressure NIBP, body temperature TEMP, respiration RESP

    Upgradeable:

    Dual body temperature (surface temperature, cavity temperature)

    End-tidal carbon dioxide: ETCO2 (mainstream method or trace sidestream method)

    Dual (single) invasive blood pressure IBP

    Anesthetic gases (isoflurane, sevoflurane, etc.)

    Blood Pressure Module for Cardell®

     

     

     

    World-renowned animal blood pressure measurement technology, providing accurate and reliable measurements for different animals, from horses to kittens, and developing different calculation methods.

    Measuring method: Oscillographic measurement technique Measuring modes: Manual, automatic, continuous. Automatic measurement time: 1, 2, 3, 4, 5, 10, 15, 30, 60 minutes.

    Measuring time: - Small cuff: 18 seconds to reach the measuring standard

    Large cuff: 26 seconds to reach measurement standard Maximum measurement time: 150 seconds

    Initial maximum perfusion pressure: 150 mmHg; if it cannot be measured the first time, increase by about 30 mmHg each time

    Pressure detection range: 20-265 mmHg

    Heart rate monitoring range: 20-300 beats per minute Accuracy: +/-5 mmHg Activity differential protection: Yes. Comes with 7 different cuffs

    Complies with AAMI/ANSI SP10, '02. Animal types vary standards and international animal care standards

     

               

     

    Nellcor's fully digital Oximax® blood oxygen module

    Nellcor's fully digital OxiMax technology provides the most accurate animal blood oxygen monitoring. This special new technology can ensure that the animal can collect signals even when it is low perfusion and moving, especially when the animal is in a critical state, it can obtain the animal's vital signs information in time, reminding the user to take appropriate measures to ensure the safety of the animal as much as possible.

    Animal-specific blood oxygen sensors (two sizes of animal-specific clips are available) can be clipped on the tongue, tail or ear of animals. Upgradeable strap-on blood oxygen belts can be used to measure blood oxygen concentration in the animal's limbs.

    Pulse rate measurement range: 35 - 250 bpm
    Pulse rate accuracy: +/-3 bpm Blood oxygen monitoring range: 0 - 100%
    Blood oxygen accuracy: - Large animals: 70-100% +/-2 digits - Small animals: 70-95% +/-3 digits
    Probe type: Nellcor animal-specific probe

     

           

     

    Temperature and respiration monitoring

    Body temperature: Temperature display - 2 channels of body temperature (upgradeable to surface and cavity probes) Temperature probe: YSI SEries 400
    Range: 9ºC to 50ºC Accuracy: +/-1ºC from 25ºC to 40ºC Respiration: Impedance Principle
    Waveform adjustment: Manual Update time: Maximum 2 seconds

     

          

     

    Fully digital ECG module

    Standard 3-lead ECG acquisition, upgradeable to 5-lead probe. Electrosurgical unit anti-interference protection ECG waveform and expanded waveform display options ECG waveform storage recall, review any abnormal ECG Upgradeable esophageal ECG 3 sensitivity filters

    3-lead: I. II. III or 5-lead: I, II, III, aVR, aVL, aVF, V Lead wire: Standard 6-Pin AAMI

    Response Filter: Filtered: 0.5 to 35 HZ Digital Filter: 0/05 to 100 HZ Defibrillator Protection: Yes Electrosurgical Protection: Yes Accuracy: +/-1 BPM Heart Rate Average: 8 beats

    Waveform display: 4 channels Waveform speed: 12.5/25/50 mm/s Range: 15-300 BPM Equipped with animal-specific ECG flat clip, different from traditional alligator clip, to reduce damage to animal skin

    Microstream® End-Tidal CO2 Module

    Mainstream end-tidal carbon dioxide monitoring/micro-flow carbon dioxide monitoring can choose mainstream type:

     

           

     

    Calculation time: 15 seconds to start, all data within 2 minutes Measurement range: 0 to 150 mmHg, 0 to 19.7%,
    0 to 20 kPa (at 760 mmHg). Rise time: Less than 60 ms.
    CO2 resolution: 0.1 mmHg 0 to 69 mmHg.
    0.25 mmHg 70 to 150 mmHg.
    CO2 accuracy: 0 - 40 mmHg +/-2 mmHg.
    41-70 mmHg +/-5% of reading.
    71 - 1000 mmHg +/-8% of reading.
    101 - 150 mmHg +/-10% of reading

    Breathing Rate Range: 0 to 150 Breaths/Minute. Breathing Accuracy: +/-1 breath.

    Calibration: No need for daily calibration by customers, no need for CO2 collection tubes and regular soda lime calibration.

    Other accessories: (flexibly equipped according to customer requirements)

     

    Microflow CO2 Probe  Esophageal ECG Probe Invasive Blood Pressure Kit

     

     

    User list:

    Renal Transplantation Laboratory, The First Affiliated Hospital of Zhengzhou University
    Weifang Animal Husbandry and Veterinary Vocational and Technical College

     

Product inquiry

Looking forward to your message !